Folia Medica Indonesiana
ISSN 0303-7932
Vol. 50 / No. 4 / Published : 2014-10
Order : 11, and page :262 - 264
Related with : Scholar Yahoo! Bing
Original Article :
Analysis of blood sample lysis rate on hemoglobin examination results using rayto rt. 7600 auto hematology analyzer
Author :
- Devi Cynthia Dewi*1
- Adang Durachim*2
- Mahasiswa Fakultas Kedokteran
- Mahasiswa Fakultas Kedokteran
Abstract :
On internal quality assurance that is pre analytical laboratory, such as in laboratory area, particularly in hospitals, the research onthe analysis of the blood sample lysis Level Examination Results Hemoglobin Using Tool "Rayto RT.7600 Auto HematologyAnalyzer, the principle flow cytometric measurement method is the amount. To make the level of the blood sample lysis sample weretreated by adding 0.5 ml of blood plus 0.2 ml physiological saline (0.85%) for normal, 0.64% NaCl starting blood lysis, NaCl 0.43%blood rather lysis, and 0.21% NaCl lysis of blood in each blood sample treatment. After doing research, the result showed that wherenormal blood sample mean = 12.9333, on blood sample began to lyse the mean = 11.890, on blood sample averaged somewhat lysis= 10.9000, and the lysis of blood sample averaged = 9.6633 and performed after statistical test by using ANOVA test, it can be seenin the statistics with ANOVA test, where P values obtained significance value (P-value) is smaller (<) compared with the value ofalpha (α) set 0.05 (95%). From these results it can be concluded that the rate of blood sample rather lysis and lysis can not be usedfor examination of the use of Hemoglobin means that lysis of blood sample can affect the result of hemoglobin. (FMI 2014;50:262-264)
Keyword :
hemoglobin, normal blood, blood lysis,
References :
Gomez KA ,(1995) Prosedur statistik untuk penelitian pertanian - : Universitas Indonesia Press
Archive Article
Cover Media | Content |
---|---|
![]() Volume : 50 / No. : 4 / Pub. : 2014-10 |
|